中文摘要: 为提高食品中汞含量测定的准确度,以国家标准物质大米和圆白菜为样品,采用微波消解和原子荧光光度计对汞含量进行测定,考察过氧化氢量、赶酸时间、硼氢化钾浓度以及载流酸度等影响因素对测定结果的影响。研究结果表明:消解过程中加入1 mL过氧化氢,消解后不赶酸,0.1%的硼氢化钾,2%的盐酸为所筛选的最佳条件。此条件显著提高汞测量的精准度。
Abstract:In order to optimize the determination method of mercury in food,the national standard material rice and cabbage were measured by microwave digestion methody atomic fluorescence spectrometer,the influence factors of H2O2,acid rush time,potassium borohydride concentration and carrier acidity were inspect.The results shwed that adding 1mL hydrogen peroxide in the digestion process,the acid is not heated to remove,potassium borohydride 0.1%,2%hydrochloric acid was the best condition of the screening,the conditions can of mercury measurement can improve the precision significantly.
keywords: food mercury element microwave digestion microwave digestion methody atomic fluorescence spectrometer
文章编号:201806029 中图分类号: 文献标志码:
基金项目:
Author Name | Affiliation |
WANG Zhen-xu,CHENG Hai-fang,ZHAO Fang,SHI Pan-pan,WEI Fa-shan | Henan Product Quality Supervision and Inspection Institute,Zhengzhou 450000,Henan,China |
引用文本: