中文摘要: 该文建立一种测定纸质食品接触材料中双酚A和双酚S含量的高效液相色谱(high performance liquid chromatography,HPLC)法。方法分析采用 Thermal Ultimate 3000SD 液相色谱仪,ZORBAX Eclipse XDB-CN(250 mm×4.6 mm,5 μm)反相柱。流动相为甲醇/水(体积比 1∶1),等度洗脱,流速 0.8 mL/min;二极管阵列(diode array detector,DAD)检测器,检测波长280 nm和259 nm。结果表明,在1 mg/L~50 mg/L浓度范围内,双酚A和双酚S的浓度与峰面积呈现良好的线性关系,相应的线性相关系数均≥0.999 8,双酚A和双酚S的检出限和定量限分别为0.10、0.02 mg/L和0.50、0.10 mg/L。方法验证过程中做了低、中、高3个水平的添加回收,平均回收率为85.2%~103.0%,相对标准偏差(relative standard deviation,RSD)均小于 5%。
Abstract:A sensitive and rapid method was developed for detection of bisphenol A (BPA)and bisphenol S(BPS)in food packaging papers using high performance liquid chromatography (HPLC).The procedure was performed with a Thermal Ultimate 3000SD equipped with a ZORBAX Eclipse XDB-CN reversed-phase column(250 mm × 4.6 mm,5 μm)using a methanol and water(volume ratio1∶1)isocratic elution.BPA and BPS were detected with a diode array detector(DAD)set at 280 nm and 259 nm.Flow rate was 0.8 mL/min.BPA and BPS were detected with good linearity within the concentration range of 1 mg/L-50 mg/L.The corresponding linear correlation coefficients were all equal to or greater than 0.999 8.Limits of detection(LOD)and limits of quantification and(LOQ)for BPA and BPS were 0.10 mg/L and 0.02 mg/L,and 0.50 mg/L and 0.10 mg/L respectively.Recovery experiments were carried out at 3 levels using low,medium,and high concentrations.Average recoveries were 85.2%-103.0% and relative standard deviations(RSD)were all below 5%.
keywords: food packaging papers bisphenol A bisphenol S high performance liquid chromatography(HPLC) isocratic elution
文章编号:202117022 中图分类号: 文献标志码:
基金项目:国家重点研发计划(2019YFC0810902、2018YFC0809104、2018YFC0809101);天津海关科研项目(2020THK011)
引用文本: